Custom
Medical Power
Supplies |
Design Examples:
|
|
CMD125
- 110W
|
|
-
130W
quad output
-
Auto-Ranging
Universal Input
-
50ms
hold-up time at full load
-
LED
Fault Indicators
-
Short
Circuit protection with auto-recovery
-
Leakage
current < 75uA at 132VAC
-
Power
fail, dc good, remote on/off.
-
Reverse
connection protection
-
4.5"W
x 8.5"L x 2.5"H (enclosed)
-
EMI
Compliance to CISPR 22, FCC Class B
-
Application:
Ultrasound P/S
|
|
|
CMD112
- 275W
|
|
-
275W
six output
-
Auto-Ranging
Universal Input w/IEC320 receptacle
-
50ms
hold-up time at full load
-
Fiber
Optic Light assembly
-
LED
Fault Indicators with AC ON/OFF switch
-
Short
Circuit protection with auto-recovery
-
Leakage
current < 50uA at 132VAC
-
Power
fail, Current limit and OVP protection
-
Remote
sense, Remote on/off, DC-OK, Reverse connection protection
-
EMI
Compliance to CISPR 22, FCC Class B
-
Application:
Ophthalmic
Surgery P/S
|
|
|
CMD116
- 300W
|
|
-
300W
five output (V5 85VDC)
-
Auto-Ranging
Universal Input w/IEC320 receptacle
-
50ms
hold-up time at full load
-
Fiber
Optic Light assembly
-
LED
Fault Indicators with AC ON/OFF switch
-
Short
Circuit protection with auto-recovery
-
Leakage
current < 50uA at 132VAC
-
Remote
sense, Remote on/off, DC-OK, Reverse connection protection
-
EMI
Compliance to CISPR 22, FCC Class B
-
Application:
Ophthalmic
Surgery P/S
|
|
|
CMD104
Series
|
|
-
300W
quad output (Four high current outputs)
-
Auto-Ranging
Universal Input
-
50ms
hold-up time at full load
-
Short
Circuit protection with auto-recovery
-
Leakage
current < 70uA at 132VAC
-
Remote
sense, Remote on/off, DC-OK, Reverse connection protection
-
EMI
Compliance to CISPR 22, FCC Class B
-
Application:
Ophthalmic
Diagnostics P/S
|
|
CMD122
Series
|
|
-
450W
six output
-
Six
high current outputs (V6 60V@1.5A)
-
Auto-Ranging
Universal Input
-
50ms
hold-up time at full load
-
Short
Circuit protection with auto-recovery
-
Leakage
current < 60uA at 264VAC
-
Remote
sense, Remote on/off, DC-OK, Reverse connection protection
-
EMI
Compliance to CISPR 22, FCC Class B
-
Application:
Surgical
Laser Supply
|
|
|
CMD105
Series
|
|
-
950W
nine output (+ 105VDC outputs)
-
Six
high current outputs (V1 2V@70A)
-
50ms
hold-up time at full load
-
PF
> 0.99 typ.
-
Short
Circuit protection with auto-recovery
-
Leakage
current < 60uA at 264VAC
-
Remote
sense, Remote on/off, DC-OK, Reverse connection protection
-
EMI
Compliance to CISPR 22, FCC Class B
-
Application:
Ultrasound
Supply
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|